Breathing Machine can adjust the treat pressure automatically
1.Dimension:220x194x112mm, 313x194x112mm
2.Weight:1.6kg,2.4kg
Breathing Machine:
Device Size
Dimensions: 220 x 194 x 112 mm, 313 x 194 x 112 mm (with InH2TM heated humidifier)
Weight: 1.6kg, 2.4 kg(with InH2TM heated humidifier)
Product Use, Transport and Storage
Operation Transport and Storage
Temperature: 5 to 30°c -20 to 55°c
Humidity: ≤ 80% Non-condensing ≤ 93% Non-condensing
Atmospheric Pressure: 860 to 1060 hPa 500 to 1060 hPa
Standard Compliance
IEC 60601-1 General Requirements for Safety of Medical Electrical Equipment
IEC 60601-1-2 Electromagnetic Compatibility
ISO 17510-1 Sleep Apnea Breathing Therapy
Mode of Operation
Continuous
AC Power Consumption
100 – 240V AC, 50/60Hz , 1.0A max
Type of Protection Against Electric Shock
Class II Equipment
Degree of Protection Against Electric Shock
Type BF Applied Part
Degree of Protection Against Ingress of Water
IPX1 –Drip-Proof, Vertical
Pressure Range
2 to 20 hPa (in 0.5 hPa increments)
Pressure Stability
2 to 20 hPa (±1 hPa)
Altitude Compensation
Manually setting, level 0, 1, 2
Automatic altitude compensation
Sound Pressure Level
<30 dB,when the device is working at the pressure of 10 hPa