Welcome to B2BAGE
Thermomagnetic oxygen analyze
Thermomagnetic oxygen analyze

Minimum Order: 0
Quality/Safety Certification: 0.00


used for continuous analyzing oxygen concentration automatically

Thermo magnetic oxygen analyzer ,       used for continuous analyzing oxygen concentration automatically, It not only applied in the  power plant,cement workshop, chemical plant ,metallurgy  etc  whereas the process control system,  also applied as monitoring instrument for  the HBOC JRC-1020 in hospital.

Working Theory

The Oxygen has high paramagnetism, the oxygen has much higher magnetic susceptibility than other gas, so in the mixed gas, the magnetic susceptibility is subject to oxygen concentration, the mixed magnetic susceptibility show the oxygen concentration among them. ,  

Sheet 1 relative magnetic susceptibility (O2as100%)

RMS

+100

+36.2

+6.16

-0.11

-04

-0.4

-0.57

-0.57

-0.57

-0.68

Gas

O2

NO

NO2

H2

N2

Steam

CO2

NH3

Ar

CH4

Technical Specifications

Detection Range and Accuracy Sheet,

At 20±5°C,The gas pressure and flow are at rated level

Sheet 2

Detection Range %

0-2.5

0-5

0-10

0-25

0-100

General Error

±2.5

±2.5

±2.5

±2.5

±2.5

1.  Repeatability:≤1%

2.  Zero Drift:≤±2.5%F.S/24h

3.  Full Scale Drift:≤±2.5%F.S/24h

4.  Linearity :≤±2.5%F.S

5.  Response Time:T90≤45s

6.  Power:220±22V  50Hz

7.  Output:0-2V(LED show % )or 0-20mA or 4-20mA(pointer meter show%) One option only

8.  The Gas should meet the following conditions:

(1)  Humidity:less than85%

(2)  Dust:≤0.01g/m3

(3)  Causticity:(SO2,H2S,NH3……)less than0.005%

(4)  Temperature: +5-50 C

(5)  Flow: 0.5-1L/min

 