RELT-100 electrochemiluminescence detecting instrument can be used in the medical test of hormones, tumor markers, and antigens.

RELT—100 flow-injection electrochemiluminescence analysis instrument

   RELT—100 electrochemiluminescence detecting instrument can be used in the medical test of hormones, tumor markers, and infectious agents. It can also be used in the following fields: food safety, environment monitoring, life science and drug development, etc.

The sampling of this detecting instrument uses high-precision injection pump, which can precisely control the doses of the sample and the carrier fluid. It can mix the sample and the carrier fluid under any proportion and inject the mixture to the surface of the electrode at required speed.

This detecting instrument is integrated with multi testing interfaces, multi analysis parameters, and multi control parts. It can indicate and analyze chemiluminescence and electrochemiluminescence signals precisely.

 

Technical indications

1 Injection pump sampling system

* Suitable injector        50uL-25mL

* Control accuracy        30%-100% of measurement range - accuracy±0.5%

* Repetitive accuracy     ≤±0.5%

* Accuracy               ≥99%(Mainly affected by the accuracy of injector)

* Control resolution       0.03% of the volume of injector (1/3000)

* Speed of injector      1.2s-20m for the whole measurement range, adjustable
2 Injection electrochemiluminescence detector
* Injection electrochemiluminescence detecting system can perform automatic flow-injection electrochemiluminescence analysis, and it is equipped with high accuracy end window photomultiplier tube.
* Programmable high accuracy negative high voltage power source and high accuracy preamplifier
* Negative high voltage range  -100~-1100V  Stability is better than 0.05%
3 High accuracy data acquisition system can perform the following functions controlled by a main computer
* Automatic/manual zero adjusting
* Gain control  1x 2x 4x 8x 16x
* Filter cut-off frequency adjustment  10~100 Hz
* Sampling rate  1~100 times/S
* The external signal input interface can receive analog signal from other instruments, inputting range: 0~5V
4 Sample detection  Luminol-H2O2-Cr(III)system
* Concentration detection range  1×10E-11~1×10E-5(g/mL)
* Detection limit  1.5×10E-12g/mL
* RSD<3%(The request for trace analysis is RSD<5%, average value of RSD=S/x) 