Type: Pathological Analysis Equipments
Place of origin: Shanghai, China(Mainland)
Format: Strip and Cassette
High sensitivity
Indicators of product technical characteristics:
1. Minimum detectable quantity: ≥ 25ng/ml
2. Sensitivity and specificity: No cross reaction occurs in the detection of AIDS-, hepatitis C-, HBsAg-, Chlamydia- and syphilis-positive samples.
Technical indicators:
Statistical results of Egens reagent on 310 samples
Egens reagent | Control reagent kit | Total | |
Positive | Negative | ||
Positive | 88 | 2 | 90 |
Negative | 1 | 219 | 220 |
Total | 89 | 221 | 310 |
Sensitivity: Se=98.88% Specificity: Sp=99.10% False positive rate =0.90%
False negative rate =1.12% Total accordance rate = 99.03%
The ond-step Alpha Fetoprotein (AFP) Test is available in strip and cassette format which conclude 2.5mm-4.0mm in strip and 3.0mm-4.0mm in cassette.
Welcome to contact me to negotiate the prices separately.