Welcome to B2BAGE
C66T Single-channel syringe pump
C66T Single-channel syringe pump

Minimum Order: 0
Quality/Safety Certification: 0.00


C66T Single-channel syringe pump

Specification:

PERFORMANCE  PARAMETER  

Flow rate

0.1ml/h --1200ml/h                                      50ml syringe

0.1ml/h --600ml/h                                       30ml syringe

0.1ml/h 399.9ml/h                                      20ml syringe

0.1ml/h 300.0ml/h                                      10ml syringe

(0.1ml/h increments, above 1000ml/h 1ml increments)

Purge

1200ml/h (50ml syringe)

600.0ml/h (30ml syringe)

399.9ml/h (20ml syringe)

300.0ml/h (10ml syringe)

Occlusion pressure

800mmHg±200mmHg(H)(106.7kpa±26.7kpa)

500mmHg±100mmHg (C)(66.7kpa±13.3kpa)

300mmHg±100mmHg (L)(40.7kpa±13.3kpa)

Precision

≤±2% (Drive accuracy≤±1%)

Total volume

0.1-9999ml

Preset limited volume

0.1-9999ml

Alarm

START reminder; near empty(NEA.EMPTY); empty; syringe disengaged; syringe dislocated; internal malfunction; occlusion; output volume equals to preset volume; battery under-voltage; battery exhausted; AC power failure; repeat alarm.

Power

AC230V, 50/60HZ (European Union) 

Internal Battery: 12V±2V

The pump can run continuously over 3 hours @5ml/h under normal conditions using internal battery after it is fully charged.

Operating condition

Temperature: +5+40

Relative Humidity: 20%90%

Transportation and storage conditions:

Temperature: -40+55

Relative Humidity: ≤95 %

Classification

Class

type CF

IP×4 ,belongs to splash-proof equipment

External dimension

307(L)×135(W)×128(H)mm

N.W.

2.2 KG

Syringe

Syringe size: 10ml/20ml/30ml/50ml

Syringe type:13kinds of disposable syringe brands can be used