Ultrasonic Transducer for Bone Density Testing
Ultrasonic Transducer for Bone Density Testing

Minimum Order: 0
Quality/Safety Certification: 0.00


Ultrasonic Transducer for Bone Density Testing
1.test bone density
2.

1.  Application

1.1Design to test Bone Density on the arm.

1.2Frequency : 1.25Mhz

 

2.  Structure

2.1  Appearance: Customized, The design can be provided by clients.

2.2  Connection wire: wire Diagram provided by clients.

2.3  Circuit Schematics: The sensor works without extra inductance & capacitance

2.4  Structures (inside):

Array element: 4pcsEffective Aperture length: 10mmWidth: 3.75mmPZT launch side polarization mark: Positive

3. Test Waveform Diagram

  

Item

Parameters 

Center frequency(Khz)

1.2475M

Overall sensitivity

-36.5db min

Receive Frequency(Khz)

1.25M

6db pulse-length

10.2us

Bandwidth

-6db 32.3%

Peak Frequency(Khz)

1.248M

Remark:  G1.25-410.3.75       Testing date: 2011.03.19 

 

Ref:

Item

Parameters

Test scope

Center frequency 

-6dB

1.25MHz

probe

Relative Bandwidth

-6dB

37.2%

probe

Pulse-length 

-20db Average

8.3us

probe

Sensitivity 

Average

-36.5dB

probe

Deviation

≤0.5 dB

Electric capacitance

Average

0.68nf

chip

Deviation

≤ 0.02 nf

Crosstalk 

≤-22dB

chip

Insulating Strength 

AC 4000V,1mA above, cut down and keep1 minute pass

probe

 

4.    Test Condition

Indoor Temperature

25±5°C

Water Temperature

25.5±0.5°C

Reflector

Stainless steel (flat)

Reflection distance

30mm

Pulse transmitter

Item No.: Agilent 33250A

Setting condition: Frequency 1.25MHz, range 10V incentive single pulse

Oscilloscopes

Agilent 54622A

 

5. Environmental Requirement (GB9706.1-2007)

 

5.1 Working conditions:

 

Environment Temperature

+0°C ~ +50°C

Relative Humidity

30%~75%

Atmospheric pressure

700Pa~ 1060hPa

5.2 Storage and transport conditions:

 

Environment Temperature

-40°C ~ +70°C

Relative Humidity

10%~100%, include condensation

Atmospheric pressure

500Pa~ 1060hPa

 

6. Probe Reliability

 

Shell defend class IPX7 test

IEC60529: 2001

Free drop test

GB9706.1-2007

Electromagnetic compatibility

GB9706.1-2007

Cable tensile test

GB9706.1-2007

Biological safety compatibility

ISO10993-5:1999

ISO10993-10:2002

Probe surface temperature

IEC60601-2-37: 2001

 

Thermal Shock Test

Test in high temperature 60°C for 30 minutes:

Test in low temperature -10°C 30 minutes;

keep testing thermal cycling for 20times

Immerse the probes into
ultrasonic couplant,
2% strong glutaraldehyde,
and 5% salt water for testing

Immerse into the reagents
(ultrasonic couplant, 2% strong glutaraldehyde, and 5% salt water) in correct order, for two hours, air dry for 2 hours, 20 times each

 

 

 