CO2 Gas Detector
1. NDIR principle
2. good stability and sensitivity
3.high precision
CO2 Gas Detector
1.The measurement principle of (NDIR)" Non-dispersive Dual-wavelength Infrared Spectrum" is adopted for PGas-31 gas infrared quantitative detector.
2.There is no radiotive emitter and it is convenient to carry; the screen of hand-held probe uses OLED,
3.it is selfluminous with high brightness, with no visual angle problem; high precision, good stability and sensitivity, and the resolution of leakage detection can reach 1ppm;
4.the response time is fast; with no consumable parts, life time of the sensor is extra long; it can detect gases include C2H2,CH4,C4H10,C2H4,C2H6, C3H8,CO2,CO and SF6, etc.
Technical parameters
Measurement principle | Infrared (NDIR) |
Measurement range | Refer to lectotype table |
Sampling mode | Pumping type |
Service temperature | -10~60°C |
Power supply | 7.2V DC, internal battery power supply |
Response time (T90s) | 1~5 second(s) |
Warm-up time | 100 seconds |
Probe unit rod length | 22cm |
Service life of sensor | >10 years |
Calibration time | Calibration once every six months |
Alarm mode | Buzzer |
Technical parameters
Parameter | Characteristics |
Measurement principle | Infrared (NDIR) |
Measurement range | Refer to lectotype table |
Sampling mode | Pumping type |
Service temperature | -10~60°C |
Power supply | 7.2V DC, internal battery power supply |
Response time (T90s) | 1~5 second(s) |
Warm-up time | 100 seconds |
Probe unit rod length | 22cm |
Service life of sensor | 10 years |
Calibration time | Calibration once every six months |
Alarm mode | Buzzer |
Lectotype table for detected gas (part gas)
Model | Measured gas | Measurement range (volume content) | Measurement precision |
PGas-31-CO2-1 | CO2 gas | 0-5000ppm | ±1%F.S. |
PGas-31-CO2-2 | CO2 gas | 0-2% | ±1%F.S. |
PGas-31-CO2-3 | CO2 gas | 0-20% | ±1%F.S. |
PGas-31-CO2-4 | CO2 gas | 0-100% | ±1%F.S. |
PGas-31-CO-1 | CO gas | 0-20000ppm | ±2%F.S. |
PGas-31-CO-2 | CO gas | 0-10% | ±1%F.S. |
PGas-31-SF6 | SF6 gas | 0-1000ppm | ±1%F.S. |
PGas-31-CH4-1 | CH4 gas (methane) | 0-4.4%(0-100%LEL) | ±1%F.S. |
PGas-31-CH4-2 | CH4 gas (methane) | 0-100% | ±1%F.S. |
PGas-31-C2H6 | C2H6 gas (ethane) | 0-2.4 %(0-100%LEL) | ±1%F.S. |
PGas-31-C3H8 | C3H8 gas (propane) | 0-1.7%(0-100%LEL) | ±1%F.S. |
PGas-31- C4H10 | C4H10 gas (butane) | 0-1.4%(0-100%LEL) | ±1%F.S. |
PGas-31- C2H2 | C2H2 gas (acetylene) | 0-2.3%(0-100%LEL) | ±1%F.S. |
PGas-31- C2H4 | C2H4 gas (ethylene) | 0-2.4%(0-100%LEL) | ±2%F.S. |