Switchable spectral bandwidth
a good many quantification and qualification functions
stability and sensitivity
UV-vis Spectrophotometer SP-2102UV/2102UVPC (UV-Vis)
--Switchable spectral bandwidth especially suits for the analysis in drug control, pharmaceuties, and scientific research, and soon.
--SP-2102UVPC is a new generation of UV/Vis Spectrophotometer with spectrum scanning function.
--This model not only reaches the instrument technical index prescribed in Drug Codex, but also has a good many quantification and qualification functions, such as wavelength scanning, single wavelength, double wavelength, triple wavelength, or multiple wavelength measurement, as well as derivative spectrum analysis, four basic calculations of spectra.
--Excellent instrument stability and sensitivity render itself to be the top one among the same class instruments.
--The performance of SP-2102UVPC competes that of double beams spectrophotometers.
Introduction:
--Switchable spectral bandwidth especially suits for the analysis in drug control, pharmaceutics, and scientific research, and so on.
--SP-2102UVPC spectrophotometer made by “Shanghai spectrum instrument Co.ltd.” has been developed on the basis of SP-2102UV, which utilizes the abundant resources and functions of MS Windows platform. SP-2102UVPC is a new generation of UV/Vis Spectrophotometer with spectrum scanning function. This model not only reaches the instrument technical index prescribed in Drug Codex, but also has a good many quantification and qualification functions, such as wavelength scanning, single wavelength, double wavelength, triple wavelength, or multiple wavelength measurement, as well as derivative spectrum analysis, four basic calculations of spectra. Excellent instrument stability and sensitivity render itself to be the top one among the same class instruments. The performance of SP-2102UVPC competes that of double beams spectrophotometers.
Parameter:
ITEM MODEL | SP-2102UV | SP-2102UVPC |
Optic System | Littrow type optic system with 1200 grooves/mm diffraction grating monochromator | |
Light Path | 100mm | 100mm |
Spectrum Bandwidth | 2nm | 2nm(or 1nm) |
Wavelength Range | 190-1000nm | 190-1100nm |
Wavelength Accuracy | ±1nm | ±0.5nm |
Wavelength Repeatability | 0.5nm | 0.2nm |
Stray Light | ≤0.1%T (at 220nm & 360nm) | |
Photometric Mode | T, A, and C measurement | |
Photometric Range | Photometric Range | |
Photometric Accuracy | ±0.3%T | ±0.3%T |
Photometric Repeatability | 0.15%T | 0.15%T |
Stability | ±0.001A/hr (at 500nm, after warming up for 1 hour) | |
Baseline Flatness | — | ±0.002A |