Ozone aging chamber
1.Ozone concentration: 0~1000 pphm
2.Temperature: measurement range RT~+55C
3.competitive price

Ozone aging chamber

 

1. Use

It is used to test rubber and rubber products ozone proof functions, and applied to: JIS K 6259ASTM1149 ISO1431 GB/T7762GB/T13642 etc.

 

2. Basic Principle

 Ozone in air takes up a little proportion, but it is the main reason for crack of rubber, imitates and strength the ozone in air, investigate the rule of ozone to rubber and identify, evaluate rubber ozone proof function, then take measures to improve rubber product.

 

3. Main features 

It is applied to international standard and keep safety of persons.

During1000pphm ozone concentration can be set at random

Speed electric machine drive sample frame configuration( satisfy static stanching test, test products running track is fit for cycloid motion theory)

Ozone concentration auto digital display controlling instrument.(touch screen type)

Double light beam QUA out of dispersion ultra ozone detection instrument

Digital display measurement on line automatically.

Measurement adjustment interface

Ozone concentration/temperature and humidity set value controlling PID auto setting

Security gate lock controlling system

 

4.Shell static painting/internal chamber is stainless steel

Gas flow indicator

Silent discharge ozone generator, with "QUA" none-dispersion ultra ozone detector

Gas purifier and thermostatic bath

Activated carbon adsorption and silica gel drying

Airflow of ozone with test samples parallel

Stabilized voltage supply inside

 

5.. Security protection configuration

Over power, circuit breaker protection, control loop over load

Electrical machine over load protection, ground protection

Over temperature protection

Alarming reminder

6.. Standard and Test methods

Satisfy dumbbell pieces and real work pieces testing samples frame GB/T7762-2003

Above static stanching test sample frame satisfy to cycloid motion theory GB/T7762-2003

Above dynamic stanching test sample frame applied to GB/T13642

 

Model

JQ-100

JQ-150

JQ-225

JQ-408

JQ-500

JQ-010

Work size(mm)

450*450*

500

500*500*

600

550*550*

720

700*700*

820

800*800*

800

1000*1000*

1000

Temperature range

     RT+10-55  degree

Temperature uniformity

      2 degree

Degree of temperature fluctuation

      0.5 degree

Ozone concentration

    10PPHm-1000PPHm

Imitation of ozone concentration

     10%

Rack rotary speed

     20-28mm/s

Stretching frequency

      0.5 Hz

  