1. Double-channels Syringe pump
2. Double CPU
3. Body weight mode, Time mode, Rate mode and link mode
4. Good quality
Maximum injection rate | 1500 ml/h (Different syringes have different maximum flow rate.) |
Injection flow rate range | 5 ml injector: 0.1~100ml/h 10 ml injector: 0.1~200 ml/h 20 ml injector: 0.1~400 ml/h 30 ml injector: 0.1~600 ml/h 50 ml injector: 0.1~1500 ml/h |
Injection increment | 0.1 |
Bolus rate | 5 ml injector: 100ml/h 10 ml injector: 200 ml/h 20 ml injector: 400 ml/h 30 ml injector: adjustable from 400~600 ml/h 50 ml injector: adjustable from 400~1500 ml/h |
Injection mode selection | 1. Rate mode; 2. Time mode; 3. Body Weight mode; 4. Link mode |
Rate mode | Rate: 0.1~1500.0ml/h (maximum value depends on the injector specifications) |
Time mode | Time: 1minute~2000 minutes Volume limit: 0.1ml~999.9 ml |
Body Weight mode | Weight: 0.1~300.0 kg Drug: 0.1~999.9 mg Volume: 0.1~999.9 ml Dose: 0.1~9999.9 (restricted to unit, drug, solution, weight and injector specification). Unit: mg/kg/h, ug/kg/min |
KVO rate | Adjustable range: 0~5.0ml/h (begin the KVO function under the occlusion situation); Setting value as 0 to cancel the KVO function when it is unnecessary in use. |
Water-proof grade | IPX1 |
Injection pressure | The maximum pressure is 300 kpa, occlusion alarm pressure gate value is 40kPa~160kPa, when pressure value is minimum, injection rate is 5ml/h, the longest alarm time is 10 minutes. |
Preset volume scope | 0.1~999.9 ml |
Preset time scope | 0~2000 minutes |
Accumulated injection volume display | 0.1~999.9 ml |
Injection accuracy | ±5% |
Power supply | 100V-240V~, 50Hz±1Hz, 30VA |
Battery | Rechargeable lithium polymer battery, 7.4V, 1600mAh |
Maximum power consumption | 10VA, running more than 2 hours at the rate of 5ml/h after being fully charged. |
Battery charge | When double-channel syringe pump is connected to AC power and the power is on, the battery will charge automatically. (about 8~14 hours to recharge fully) |
Fuses | T 2AL 250V~ |
Displayed and indicated information | Injection rate, accumulated volume, syringe specification, battery capacity, bed No., AC power indicator, operation indication, etc. |
State indication information | Stop, injection, bolus, KVO ( If KVO rate is set as 0, it doesn’t display KVO when occlusion alarm occurs) |
Alarm functions | Near completion, Over, occlusion, disengage, low battery, abnormal 1, abnormal 2, abnormal 3, abnormal 4, serious low battery |
Size of the outer shell | 320 mm×110 mm×200mm (length×width×height) |
Net weight (excluding power cord) | 3.4 kg |
Classification | ClassII, Type BF |
Outer shell material | ABS Plastic |
Operating condition | Environment temperature:+5°C~+40°C, air pressure: (86~106)kPa, relative humidity: ≤80% |
Storage condition | Environment temperature:-40°C~+55°C, air pressure: (50~106)kPa, relative humidity: ≤95%. No poisonous gas and well ventilation |
Applicable syringes | National standard syringes: 5ml, 10 ml, 20 ml, 30ml, 50 ml |
Standards | GB9706.1-2007, GB 9706.27-2005, ISO7886-1:1993, GB9969.1-1998, GB/T14710-1993, GB191-2008, YY0466-2003 |