Minimum Order: 0
Quality/Safety Certification: 0.00
Dewpoint Transmitter DMT142
Dewpoint Transmitter DMT142 Dewpoint temperature measurement |
| Standard version Measurement range Accuracy (-50 ... +60 °C / -58 ... +140 °F) Extended version Measurement range Accuracy (-60 ... +60 °C / -76 ... +140 °F) Sensor | -60 ... +60 °C (-76 ... +140 °F) Td ± 3 °C (± 5.4 °F) Td
-80 ... +60 °C (-112 ... +140 °F) Td ± 3 °C (± 5.4 °F) Td
Vaisala DRYCAP® 180D
|
ppmvolume concentration measurement |
| Measurement range Accuracy at +20 °C (+68 °F), 1013 mbar | 70 ... 200 000 ppm 7.3 ppm + 9.2 % of reading |
Operating environment |
| Temperature Relative humidity Pressure | 0 ...+60 °C (+32 ... +140 °F) 0 ... 100 %RH 0 ... 20 bara(290 psia) |
Inputs and outputs |
| Operating voltage with voltage output with current output Analog output (scalable) | 12 ... 28 VDC 18 ... 28 VDC 4 ... 20 mA (3-wire) 0 ... 1 V 0 ... 5 V
|
Mechanics |
| Weight Material (housing) Housing classification | 118 g (4.16 oz) stainless steel body (AISI 316L), plastic cap (ABS/PC) IP65 (NEMA 4) |