CS dew point sensor FA 300 Ex
CS dew point sensor FA 300 Ex

Minimum Order: 0
Quality/Safety Certification: 0.00


FA 300 Ex may only be used in connection with approved sensor feeding instruments or safety barriers or galvanic separating

The CS dew point sensor FA 300 Ex provides reliable and long term stable dew point monitoring in applications where explosive gases can occur from time to time or permanently. It’s approved for use in explosive areas of  zone 1. II 2G EEx ia IIC T4

FA 300 Ex may only be used in connection with approved sensor feeding instruments or safety barriers or galvanic separating elements with max.:

U0 = 30 V max.
I0  = 100 mA max.
P0  = 1 W max.

The measured dew point is output via a 4-20 mA signal output (loop powered).

Technical data FA 300 Ex
   Measuring range
          Dew point
          Relative humidity
          Temperature

-80...+20 ºC                
  0...99.9%                   
-30...+70 ºC  
   Accuracy±2 ºC dew point @ -40 ºC
   Pressure range-0.1 ... 30 MPa
   Power supply16 ... 30 VDC
   Ambient conditions0 ... 50 ºC
   Response time t90< 30 sec (descending)
< 10 sec (ascending)
   Output signal4...20 mA, 2-wire
   ConnectionDIN 43650, 4 pole
   CasingPC
   ClassificationIP 65
   Protection classII 2G EEx ia IIC T4
   Process connectionG 1/2" thread (ISO 228/1)
   Transport temperature-30 ... +70 ºC display

 

 